PE‐HVT‐AM series is an automatic HIPOT testers which are build to test the breakdown voltage of insulation dielectric, while ensuring user safety. These testers are completely automatic in operation with integrated matrix printer. The tester comes in variety of options in AC, DC, AC&DC, and in a variety of voltage ratings (0‐500kV) and current ratings (0‐500mA) levels, to cater to different requirements of customers
Power Supply- Single Phase AC 200V‐250V, or Single Phase AC 90V ‐150V
Voltage output- Continuously variable 0‐FSV (steps of 0.01% of FSV)
Display Voltage accuracy- 1% + 2 digits (“True HV-HIPOT test”)
Trip Current- programmable from 10% to 100% of FSV current, in steps of 1%
Current Display- 100pA to FSV
Display Current accuracy- 1% + 10 digits
Insulation Resistance- 10MΩ to 10TΩ
Size- 450mm(W)×310mm(D)×360mm(H)
Gross Weight- 50kgs (including accessories), except packing
Stand time- 0 to 10 mins
Test time- 0 to 10 mins
Break Sequence- 10 break sequence (eg. go to 10kV>hold for 5 mins>go to 20kV>hold for 4 mins>go to 30kV>hold for 3 mins>go to 40kV>hold for 2 mins>go to 50kV>hold for 1 min>end test)
Voltage Ramp‐up- 1kV/sec to 5kV/sec (variable in steps of 0.5kV/sec)
Sequence Repetition- 1 to 10 times
Interlock for Zero start
Auto trip and reset
Operating Temp- -10°C to +60°C
Humidity- < 90% non‐condensing
Max Voltage 0‐500kV (continuously variable)
Selectable AC o/p, DC o/p. DC polarity reversal possible
Programmable current trip levels 10% to 100% of maximum
World’s only “True HV-HIPOT set”…measures very HV secondary voltage across output terminal, instead of across HV primary
Complete Automatic operation. Just select and Start!
“Most accurate test set”. 1% Accuracy over entire test range.
Very high Resolution 0.01% of FSV of voltage
Leakage current measurement from 100pA to FSV of current
Insulation Resistance (IR) value measurement
Polarization Index (PI) test facility
Programmable test time 0‐99 mins
Programmable voltage ramp rate from 1kV/sec to 5kV/sec
Proof/Voltage withstand test possible with 10 break sequences
Repetitive test sequence (up to 10 iterations) and average readings (from first to last), min, max and standard deviation.
Safety features like ground check, zero start interlock and automatic reset incorporated for user safety
User and object information included
Touch panel, Backlit color LCD screen for comfortable user interface
Integrated thermal Printer
Internal memory for 100 test data storage and recall via USB port
Programmable 25 custom test sequences as per requirement
Remote user terminal available optionally, with operating distance range of 50m
High Voltage 500kV digital meter (PE‐HVM‐500) 1% accuracy available optionally for on‐site calibration
Testing high voltage insulation of various electrical object such as transformer bushing & switchgear equipment etc.
High Voltage Insulated test probes – 1 set
Crocodile pins – 2 nos
Connecting AC mains cable – 1 no
Printer roll – 1 no (already loaded)
User manual – 1 no
Test certificate – 1 no
Warranty Certificate – 1no (Standard warranty 1 yr)
Fill out the form below to download datasheet & we will contact you soon
Fill out the form below and we will contact you as soon as possible!