PE-ACDF-12/200 is a fully automatic capacitance & tan-delta test system with inbuilt 12 kV AC power source measurement bridge and a reference SF6 gas filled standard capacitor.
The kit is simple and easy to operate with a large backlit LCD display, and soft touch membrane buttons. The kit is extremely light in weight & portable with a USB serial output port for users to download test results. The kit also has internal storage capacity of up to 99 tests data. Integrated thermal Printer of kit helps in printing results on site.
The instrument is calibrated in our NABL accredited test laboratory which delivers results traceable to national standards.
Input Supply- 1-phase 220V 50Hz, or, 110V 60Hz (also available on request)
Internal HV frequency- 45 to 55 Hz (single) 45/55Hz (auto multiple frequency measure & average for interference suppression)
Dissipation Factor Range- no limit
DF Resolution- 0.001% (1 X 10-5)
DF Accuracy- 1% + 4 digits
Capacitance Range- 3pF to 50000pF @ 12kV 60pF to 1uF @ 0.5kV
Capacitance Range Accuracy- 1% + 1 digit
Test Voltage range- 0.5kV to 12kV, 200mA(max)
Test time- 30s typical
Test Voltage accuracy- 1% + 10 digits
CVT test mode- O/p Voltage 3 to 50V, O/p Current 3 to 30A
Humidity- <90% condensing
Dimensions (excluding accessories)- 45 X 35 X 35 (all dims. in cm)
Test Voltage resolution- 1V
Input Current Range- 10uA to 5A
Printer- Thermal printer
Communication Port- RS-232
Operating Temperature- –10°C to 50°C
Weight including accessories- 35 kg
Storage Temperature- -20°C to 60°C
Just set test parameters, push start, and everything is automatically done in sequence right up to printing of test results.
High Interference suppression, good signal-to-noise ratio.
UST, GST ad GSTg test modes
CVT test facility
Input overvoltage protection
Output short-circuit protection
Can be used with external power supply up to 12kV for enlarged current to 1A, for generator related testing
Electronic Calibration done at our lab, with results traceable to national standards.
RS-232 output
Poor ground rejection
Fault protection and indication
Thermal Printer integrated
Transformer testing
Testing Bushings, CTs, PTs, etc. at substations
Testing Rotating machines (LV Motors/Generators) on-site
Test leads (HV, LV, Gnd.) – 1 set (20mtrs length)
Connecting AC mains cable – 1 no
Printer roll – 1 no (already loaded)
User manual – 1 no
Test certificate – 1 no
Warranty Certificate – 1no (Standard warranty 1 yr)
Fill out the form below to download datasheet & we will contact you soon
Fill out the form below and we will contact you as soon as possible!